Change of name from AnaConDa to SedacondaHow do I use the Sedaconda ACD (anaesthetic conserving device from Sedana Medical) with Hamilton Medical ventilators?
The Sedaconda ACD is an anesthetic delivery system developed for the administration of volatile anesthetics such as Isoflurane or Sevoflurane to invasively ventilated patients. Use of the Sedaconda system in combination with all Hamilton Medical ventilators has been successfully validated by Sedana Medical.
The Sedaconda applicator and filter should be connected in the same way as a bacterial and viral HMEF.
Figure 1 shows the correct positioning and connection of the Sedaconda system.
The order of the adapters should be as follows:
You can also find information in your ventilator’s Operator’s Manual about connecting a bacterial and viral HMEF closer to the patient to protect the flow sensor from humidity.
The expiratory valve is connected to the scavenging system by means of an adapter (which is part of the circuit).
It should be noted that use of the Sedaconda system may slightly increase airway resistance.
Useful adaptations when using Hamilton Medical Ventilators with the Sedaconda system:
Relevant devices: All