Atrás

Use of Sedaconda with Hamilton Medical ventilators

Artículo

Autor: Thomas Reimer

Fecha: 28.11.2024

Last change: 28.11.2024

Change of name from AnaConDa to Sedaconda

How do I use the Sedaconda ACD (anaesthetic conserving device from Sedana Medical) with Hamilton Medical ventilators?

Use of Sedaconda with Hamilton Medical ventilators

Background

The Sedaconda ACD is an anesthetic delivery system developed for the administration of volatile anesthetics such as Isoflurane or Sevoflurane to invasively ventilated patients. Use of the Sedaconda system in combination with all Hamilton Medical ventilators has been successfully validated by Sedana Medical.

 

Connecting the Sedaconda

The Sedaconda applicator and filter should be connected in the same way as a bacterial and viral HMEF.

Figure 1 shows the correct positioning and connection of the Sedaconda system.

The order of the adapters should be as follows:

  • Y-piece
  • CO2 sensor
  • Flow Sensor
  • Sedaconda

You can also find information in your ventilator’s Operator’s Manual about connecting a bacterial and viral HMEF closer to the patient to protect the flow sensor from humidity.

The expiratory valve is connected to the scavenging system by means of an adapter (which is part of the circuit).

Diagram showing positioning and connection of Sedaconda system
Correct setup with the Sedaconda
Diagram showing positioning and connection of Sedaconda system
Correct setup with the Sedaconda

Helpful tips

It should be noted that use of the Sedaconda system may slightly increase airway resistance.

Useful adaptations when using Hamilton Medical Ventilators with the Sedaconda system:

  • Use pressure trigger
  • Do not activate TRC

Relevant devices: All