Current management practices and outcomes in weaning from mechanical ventilation

Article

Author: Jean-Michel Arnal, Senior Intensivist, Hopital Sainte Musse, Toulon, France

Date of first publication: 04.12.2023

A recent publication documents results from the largest international study so far to investigate the management and outcomes of weaning in ICU patients.

Current management practices and outcomes in weaning from mechanical ventilation

Takeaway messages

  • Despite the impact of weaning on patient outcomes and cost, the weaning process is still poorly understood.
  • The WEAN SAFE study is the largest international study to investigate management practices and outcomes in a real-world population and included 5,869 ICU patients from 50 countries.
  • The study confirmed that both the patient’s physiological characteristics and weaning practices contribute to weaning failure.
  • Deep sedation at the time of the separation attempt and delays between meeting the weaning eligibility criteria and the first separation attempt were both independently associated with weaning failure.

What is behind the WEAN SAFE study?

Weaning from mechanical ventilation refers to the process of liberating the patient from the ventilatory support and artificial airway. Weaning represents a significant amount of the total invasive ventilation duration in ICU patients, and is associated with both short- and long-term outcomes (Herridge MS, Azoulay É. Outcomes after Critical Illness. N Engl J Med. 2023;388(10):913-924. doi:10.1056/NEJMra21046691​), as well as cost. However, the weaning process is still poorly understood and there are no clear definitions of the beginning and end of weaning. The Worldwide Assessment of Separation of Patients from Ventilatory Assistance (WEAN SAFE) study is the largest international multicenter observational study aimed at understanding the management practices and outcomes in a real-world population of ICU patients (Pham T, Heunks L, Bellani G, et al. Weaning from mechanical ventilation in intensive care units across 50 countries (WEAN SAFE): a multicentre, prospective, observational cohort study [published correction appears in Lancet Respir Med. 2023 Mar;11(3):e25]. Lancet Respir Med. 2023;11(5):465-476. doi:10.1016/S2213-2600(22)00449-02​). The study was supported by the European Society of Intensive Care Medicine (ESICM) and endorsed by a number of national societies. The study enrolled 5,869 patients out of 10,232 patients screened in 481 ICUs from 50 different countries. Each ICU enrolled patients for four consecutive weeks between October 2017 and June 2018. 

Patient characteristics and weaning eligibility criteria

All patients over 16 years of age and mechanically ventilated in the ICU were screened, and were included if they were still receiving invasive mechanical ventilation two calendar days after intubation. The weaning eligibility criteria were FiO2 below 50%, PEEP below 10 cmH2O, no or low doses of vasopressors, and no paralyzing agent. The level of consciousness was not taken into account for assessing weaning eligibility, as it is a criterion for extubation. The start of weaning was defined as the first weaning trial, whatever the method used, or the first extubation without a weaning trial. Weaning success was defined as extubation or disconnection from the tracheostomy tube without death or reintubation within the following seven days. Patient’s characteristics and weaning outcomes are presented in Figure 1.

Diagram showing characteristics and outcomes
Figure 1: Patient characteristics and weaning outcomes
Diagram showing characteristics and outcomes
Figure 1: Patient characteristics and weaning outcomes

Time until first attempt and weaning duration

More than 90% of the included patients presented signs of spontaneous activity and met the weaning eligibility criteria at a median of three days following tracheal intubation. The median time from meeting the weaning eligibility criteria and the first attempt at separation was 1 day, but in 22% of patients this delay was prolonged to more than 5 days. Of the patients who were weaned successfully, the weaning process was short in 77% (≤ 1 day), of intermediate duration in 12% (2–6 days), and prolonged in 11% (≥ 7 days).

The demographic factors associated with a delay of more than 1 day between meeting the eligibility criteria and the first separation attempt included frailty, admission for trauma, and a non-traumatic neurological event. Critical illness severity as measured by the SOFA score was also associated with a prolonged delay. The continuous use of neuromuscular blockades and use of moderate or deep sedation on the first day of fulfilling the weaning criteria were also associated with a delayed separation attempt. 

Factors associated with failure

Factors associated with weaning failure were older age, frailty, immunocompromised status, severity of critical illness, cardiac arrest, non-traumatic neurological event, pre-existing limitation of care, and severe respiratory dysfunction. The presence of deep sedation at the time of the separation attempt, and a delay between meeting the eligibility criteria and the first separation attempt were independently associated with weaning failure. A sensitivity analysis was performed including the number of beds in the ICU and the same results were found.

What have we learned?

Weaning failure and delayed weaning are a burden for the patient, family, and healthcare system. The WEAN SAFE study confirms that both the patient’s physiological characteristics and weaning practices contribute to weaning failure. One major driver behind delays to the first separation attempt that eventually increase weaning failure is the sedation management at the time of meeting the weaning eligibility criteria. Every effort to prevent moderate and deep sedation at the time of assessing the weaning eligibility may have a positive impact, shortening the invasive mechanical ventilation duration and increasing the weaning success rate. Another lesson from the WEAN SAFE study is that a delay in performing the first weaning trial after patients have met the eligibility criteria is associated with weaning failure. Sometimes, this delay is due to the patient’s status and there is a good reason for not performing the weaning trial. On other occasions, however, the delay is due to under-recognition of the patient’s readiness-to-wean or that caregivers are not available to perform the weaning trial in a timely manner. The recent shortages of ICU nurses, respiratory therapists, and physicians are impacting on the organization in the ICU and may affect the weaning process. 

One point of discussion is whether we should wait for the patient to be awake and fully responsive to perform the first weaning trial. The authors support the approach that the first weaning trial should be performed even if the extubation criteria (consciousness, cough, etc.) are not fulfilled. By doing this, the clinician can recognize if the patient would be ready to wean and stop sedation accordingly (Pham T, Heunks L, Bellani G, Brochard L, Laffey J; WEAN SAFE Investigators. WEAN SAFE and the definition of the first separation attempt - Authors' reply. Lancet Respir Med. 2023;11(5):e44. doi:10.1016/S2213-2600(23)00089-93​). 

How can ventilators support the process?

Hamilton Medical ventilators provide some tools to support clinicians in the weaning process and, in particular, to prevent delays in recognizing the readiness-to-wean criteria and performing a weaning trial. The Vent Status panel is a simple display of the weaning eligibility criteria (see Figure 2). Even without a lot of cognitive involvement or deep knowledge, one look at the screen can tell the clinician whether that patient is ready for a weaning trial. If vasopressor doses are low, the clinician is encouraged to stop sedation and perform a weaning trial. 

For the patients ventilated with INTELLiVENT-ASV, the Quick Wean function is an automatic weaning protocol to support the caregivers in the weaning process. Quick Wean can be activated as soon as the patient recovers spontaneous breathing, whatever the level of oxygen, pressure support, or sedation. The ventilator decreases oxygen, PEEP, and pressure support within the safety rules of the mode, screens the weaning eligibility criteria, and performs a weaning trial. The weaning trial aborts automatically if the physiological variables are outside of target ranges. The clinician has to manage the sedation and decide when it is time to extubate.

Screenshot of Vent Status panel on ventilator
Figure 2: Vent Status panel showing weaning eligibility criteria
Screenshot of Vent Status panel on ventilator
Figure 2: Vent Status panel showing weaning eligibility criteria

Conclusion


In conclusion, the WEAN SAFE study is a large, international, multicenter observational study that showed important variations in weaning practices and identified some modifiable factors to decrease weaning duration and increase weaning success.

Outcomes after Critical Illness.

Herridge MS, Azoulay É. Outcomes after Critical Illness. N Engl J Med. 2023;388(10):913-924. doi:10.1056/NEJMra2104669

Weaning from mechanical ventilation in intensive care units across 50 countries (WEAN SAFE): a multicentre, prospective, observational cohort study.

Pham T, Heunks L, Bellani G, et al. Weaning from mechanical ventilation in intensive care units across 50 countries (WEAN SAFE): a multicentre, prospective, observational cohort study [published correction appears in Lancet Respir Med. 2023 Mar;11(3):e25]. Lancet Respir Med. 2023;11(5):465-476. doi:10.1016/S2213-2600(22)00449-0



BACKGROUND

Current management practices and outcomes in weaning from invasive mechanical ventilation are poorly understood. We aimed to describe the epidemiology, management, timings, risk for failure, and outcomes of weaning in patients requiring at least 2 days of invasive mechanical ventilation.

METHODS

WEAN SAFE was an international, multicentre, prospective, observational cohort study done in 481 intensive care units in 50 countries. Eligible participants were older than 16 years, admitted to a participating intensive care unit, and receiving mechanical ventilation for 2 calendar days or longer. We defined weaning initiation as the first attempt to separate a patient from the ventilator, successful weaning as no reintubation or death within 7 days of extubation, and weaning eligibility criteria based on positive end-expiratory pressure, fractional concentration of oxygen in inspired air, and vasopressors. The primary outcome was the proportion of patients successfully weaned at 90 days. Key secondary outcomes included weaning duration, timing of weaning events, factors associated with weaning delay and weaning failure, and hospital outcomes. This study is registered with ClinicalTrials.gov, NCT03255109.

FINDINGS

Between Oct 4, 2017, and June 25, 2018, 10 232 patients were screened for eligibility, of whom 5869 were enrolled. 4523 (77·1%) patients underwent at least one separation attempt and 3817 (65·0%) patients were successfully weaned from ventilation at day 90. 237 (4·0%) patients were transferred before any separation attempt, 153 (2·6%) were transferred after at least one separation attempt and not successfully weaned, and 1662 (28·3%) died while invasively ventilated. The median time from fulfilling weaning eligibility criteria to first separation attempt was 1 day (IQR 0-4), and 1013 (22·4%) patients had a delay in initiating first separation of 5 or more days. Of the 4523 (77·1%) patients with separation attempts, 2927 (64·7%) had a short wean (≤1 day), 457 (10·1%) had intermediate weaning (2-6 days), 433 (9·6%) required prolonged weaning (≥7 days), and 706 (15·6%) had weaning failure. Higher sedation scores were independently associated with delayed initiation of weaning. Delayed initiation of weaning and higher sedation scores were independently associated with weaning failure. 1742 (31·8%) of 5479 patients died in the intensive care unit and 2095 (38·3%) of 5465 patients died in hospital.

INTERPRETATION

In critically ill patients receiving at least 2 days of invasive mechanical ventilation, only 65% were weaned at 90 days. A better understanding of factors that delay the weaning process, such as delays in weaning initiation or excessive sedation levels, might improve weaning success rates.

FUNDING

European Society of Intensive Care Medicine, European Respiratory Society.

WEAN SAFE and the definition of the first separation attempt - Authors' reply.

Pham T, Heunks L, Bellani G, Brochard L, Laffey J; WEAN SAFE Investigators. WEAN SAFE and the definition of the first separation attempt - Authors' reply. Lancet Respir Med. 2023;11(5):e44. doi:10.1016/S2213-2600(23)00089-9

Related articles. Get a deeper look